Micro Imager - Operating principle |
- The Beta particle emitted by the sample (1) produces a light spot in the scintillator sheet placed on top of the sample.
- This light enters the intensifier tube (5) and is amplified by the double microchannel plate of the tube .
- Each light spot is measured by the CCD chip and individually counted or rejected.
- The actual center of the light spot is calculated to obtain maximum precisional accuracy on where the associated Beta particle entered the detector. This ensures very high resolution.
- Beta particles are counted and accumulated in real time and on the autoradiographic image displayed on the PC screen.